2012/05/29 Test 5
Test of small movements of the focuser depending of the position of the focuser. It was before we find that one screw was loosened.
Test of focuser's small movements
- Focuser horizontaly, in park position
| Order | Comparator |
| 2.00 |
2.00 |
| 2.10 |
2.10 |
| 2.20 |
2.20 |
| 2.30 |
2.31 |
| 2.40 |
2.41 |
| 2.50 |
2.52 |
| 2.60 |
2.63 |
| 2.70 |
2.73 |
| 2.80 |
2.84 |
| 2.90 |
2.94 |
| 3.00 |
3.05 |
| 2.00 |
2.00 |
| 3.00 |
3.05 |
| 3.10 |
3.15 |
| 3.20 |
3.26 |
| 3.30 |
3.37 |
| 3.40 |
3.48 |
| 3.50 |
3.58 |
| 3.60 |
3.68 |
| 3.70 |
3.80 |
| 3.80 |
3.90 |
| 3.90 |
4.01 |
| 4.00 |
4.12 |
| 3.00 |
3.05 |
| 4.00 |
4.12 |
- Focuser under
| Order | Comparator |
| 2.00 |
2.00 |
| 2.10 |
2.11 |
| 2.20 |
2.21 |
| 2.30 |
2.32 |
| 2.40 |
2.43 |
| 2.50 |
2.54 |
| 2.60 |
2.64 |
| 2.70 |
2.74 |
| 2.80 |
2.84 |
| 2.90 |
2.96 |
| 3.00 |
3.07 |
| 2.00 |
2.01 |
| 3.00 |
3.07 |
| 3.10 |
3.17 |
| 3.20 |
3.28 |
| 3.30 |
3.38 |
| 3.40 |
3.47 |
| 3.50 |
3.58 |
| 3.60 |
3.68 |
| 3.70 |
3.79 |
| 3.80 |
3.90 |
| 3.90 |
4.01 |
| 4.00 |
4.12 |
2.00
|
2.01
|
| 4.00 |
4.12 |
- Focuser above
The zero between this position and the last one, has changed by 120 microns
| Order | Comparator |
| 2.00 |
2.00 |
| 2.10 |
2.09 |
| 2.20 |
2.19 |
| 2.30 |
2.29 |
| 2.40 |
2.39 |
| 2.50 |
2.49 |
| 2.60 |
2.60 |
| 2.70 |
2.70 |
| 2.80 |
2.80 |
| 2.90 |
2.91 |
| 3.00 |
3.01 |
| 2.00 |
1.98
|
| 4.00 |
4.07 |
| 3.90 |
3.97 |
| 3.80 |
3.86 |
2.00
|
1.98
|